Products

2-zone temperature shock test chamber

Technical specifications

Main technical parameters

Mode

WBE-LR2-49LA/B/C

WBE-LR2-100LA/B/C

WBE-LR2-150LA/B/C

WBE-LR2-250LA/B/C

WBE-LR2-450LA/B/C

WBE-LR2-1000LA/B/C

Volume(L)

49L

100L

150L

250L

450L

1000L

Basket size WxDxH(mm)

400×350×350

500×500×400

600×500×500

700×600×600

800×800×700

1000×1000×1000

Outer size WxDxH(mm)

1180x1850x1913

1280X1900X2215

1380x2000X2215

1480x2175X2215

1480x2376x2410

1780x2710X2876

Basket temperature range

150℃~ABC

150℃~ABC

150℃~ABC

150℃~ABC

150℃~ABC

150℃~ABC

A -40

B -55

C-65

A -40

B -55

C-65

A -40

B -55

C-65

A -40

B -55

C-65

A -40

B -55

C-65

A -40

B -55

C-65

Overall power KW

7.75

8.75

10.75

12.5

14

15.5

21.25

24.25

27.25

28

31

35.5

37.9

45.4

45.4

50

57.5

57.5

Tonnage of water tower RT

Air cooling is not required

5

Air cooling

5

8

8

10

12

10

12

15

15

20

20

20

25

25

Water discharge m³/h

/

/

2.35

/

3.9

4.68

6.24

7.8

9.36

7.8

9.36

11.7

11.7

15.6

15.6

15.6

19.5

19.5

Pipe diameter DN/mm

/

/

40

/

40

40

50

50

50

50

50

50

50

65

65

65

65

65

Window size W*H(mm)

W210*H300*2

W280*H350*2

W280*H350*2

W290*H420*2

W350*H500*2

W350*H500*2

Noise(dB)

≤72dB

≤72dB

≤72dB

≤75dB

≤75dB

≤75dB

Heat storage temperature time

RT~200℃ approximately take≤30min

Storage temperature time

A:+20~ -50℃ approximately take 35min      B:+20~ -65℃ approximately take ≤45min       C:+20~ -75℃ approximately take 55min

High temperature impact range

+60℃~ +150℃

Low temperature impact range

A:-10℃~ -40℃            B:-10℃~ -55℃             C:-10℃~ -65℃

Temperature fluctuation

≤±0.5℃

Temperature uniformity

≤2.0℃ non-load

Temperature deviation

≤±2.0℃ non-load

Basket conversion time

5S

Temperature recovery time

≤ 5min

High temperature exposure 15min/ low temperature exposure 15min

Continuous testing without defrosting times

1000 times (without defrosting) optional if required

Operating ambient temperature

Air cooling ambient temperature requirements 5~ +30℃/ Water cooling ambient temperature requirements +5~ +40℃ (water cooling requires a water tower or ice water machine)

 

Test product mobile type, preheating zone, precooling zone two zone structure

Weight (approx) (Kg)

1000

1200

1600

1900

2500

3500

Temperature-time resolution

Temperature:0.01℃;Time: 0.01min。

Standard fittings

insulated glass viewing window 1pcs, samples rack 2pcs, measuring hole diameter 50mm*1pcs, power cord, instruction manual.

Power suply

AC380V±10%,50Hz±1Three-phase four-wire + ground cable (3/N/PE), the ground resistance is < 4Ω

Related products