It is used to detect the performance of products under rapid temperature changes and extreme temperature conditions. Mainly used in semiconductor chips, scientific research institutions, quality inspection, new energy, optoelectronic communications, aerospace , automotive industry, LCD display, medical and other science and technology industries.
Meet test criteria
1. GB/T 2423.1 Low temperature test method
2.GJB 150.3 high temperature test
3.GB/T 2423.2 high temperature test method
4.GJB 150.4 low temperature test
5. GB/T2423.34 humid heat cycle test method
6. GJB 150.9 damp heat test
7.IEC60068-2 temperature and humidity test method
8. MIL-STD-202G-103B humidity test
1.Technical parameters: |
|||||||
Name |
Rapid temperature change (humid heat) test chamber |
||||||
Model number |
WBE-KSH-150L |
WBE-KSH-225L |
WBE-KSH-512L |
WBE-KSH-1000L |
WBE-KSH-2400L |
WBE-KSH-3400L |
WBE-KSH-4600L |
Volume(L) |
150 |
225 |
512 |
1000 |
2400 |
3400 |
4600 |
Test room dimensions (WxDxH)(mm) |
500x500x600 |
600x500x750 |
800x800x800 |
1100x900x1100 |
1300x1400x1300 |
1500x1500x1500 |
1800x1500x1700 |
Window (WxH) (mm) |
210x300 |
280x350 |
360x510 |
360x510 |
360x510 |
360x510*2 |
360x510*2 |
Linear temperature range and linear temperature rate /min |
-40~125 °C (AX linear 2 °C /5 °C /10 °C /15 °C /20 °C /25 °C; AF Nonlinear 2℃/5℃/10℃/15℃/20℃/25℃) |
||||||
|
AX2、AX5、AX10、AX15、AX20、AX25 , AF2、AF5、AF10、AF15、AF20、AF25; For example, AX15: linear 15 ° C, linear range -40~125 ° C; AF15: Nonlinear (average) 15 ° C, nonlinear range: -40~125 ° C |
||||||
Nonlinear (average) temperature range and nonlinear (average) temperature rate /min |
-55~125℃ (CX linear 2℃/5℃/10℃/15℃; CF nonlinear 2℃/5℃/10℃/15℃/20℃/25℃) |
||||||
|
CX2、CX5、CX10、CX15,CF2、CF5、CF10、CF15、CF20、CF25; For example, CX10: linear 10℃, linear range -55~125℃; CF10: Nonlinear 10℃ (average), nonlinear range: -55~125℃. |
||||||
Environmental requirement |
Air cooling ambient temperature requirements +5 to +30 ° C/Water cooling ambient temperature requirements +5 to +40 ° C (Water cooling requires a water tower or water cooler). |
||||||
Temperature performance index: |
|||||||
Temperature range |
-70℃~+150℃ |
||||||
Temperature uniformity |
≤2.0℃ |
||||||
Temperature deviation |
≤±2.0℃ |
||||||
Temperature fluctuation |
≤±0.5℃ (≤±0.5℃, according to GB/T5170-1996). |
||||||
Power source |
AC380V±10%,50Hz±1 Three-phase four-wire + ground cable (3/N/PE), grounding resistance less than 4Ω. |
||||||
Humidity performance index: |
|||||||
Temperature range |
+10℃~+95℃ |
||||||
Humidity range |
10%RH to 98%RH (5% to 98%R.H; Customized for special conditions) |
||||||
Humidity uniformity |
≤3.0%RH |
||||||
Humidity deviation |
+2.0%RH ~ -3.0%RH (± 3.0%RH is expressed as GB/T 2423.3-2008) |
||||||
Humidity fluctuation |
≤±2.0%RH。 |
||||||
Temperature and humidity all test points |
For details, see 2.2 T/H Control Capability Diagram. |
||||||
Time resolution of temperature and humidity |
Temperature :0.01℃; Humidity: 0.1%; Time: 0.01min. |
||||||
Standard configuration |
observation window (double hollow tempered conductive glass) 1pcs; Test hole Φ50/Φ100mm (on the left) 1pcs; Sample rack 2 layers 1pcs; camera light (LED light strip) 1pcs in the box; small casters 4pcs; power cable 1pcs. |