Apply to test the performance of various products in heat, cold, dry and wet resistance. Mainly apply in semiconductor chips line, scientific research institutions, quality inspection, new energy, optoelectronic communications, aerospace military, automotive industry, LCD display, medical and other science and technology industries.
Test criteria
1.GB/T 2423.1 Low temperature test method
2.GB/T 2423.2 High temperature test method
3.GB/T2423.34 humid heat cycle test method
4.IEC60068-2 temperature and humidity test method
5.GJB 150.3 High temperature test
6.GJB 150.4 Low temperature test
7.GJB 150.9 Humidity heat test
8.MIL-STD-202G-103B Humidity test
一.technical parameters: |
|||
Name |
Double-door Temperature and Humidity test chamber |
||
Mold |
WBE-SDJ-1500LA |
WBE-SDJ-1500LB |
WBE-SDJ-1500LD |
Valume(L) |
1500L |
||
Inner size(WxDxH)(mm) |
1500x1000x1000 |
||
Outer size(WxDxH) (mm) |
1700x2000x2150 |
||
Window(WxH)(mm) |
290x420*2 |
||
Noice(dB) |
<70 |
||
Power supply(KW) |
12.5 |
9.5 |
15.5 |
Voltage AC(V) |
380 |
||
Weight(kg) |
950 |
850 |
1000 |
Environmental requirement |
Air cooling ambient temperature requirements +5~ +30℃/Water cooling ambient temperature requirements+5~ +40℃(Water cooling requires a water tower or chiller) |
||
Air speek |
≤1.7m/s |
||
Load |
≤600W |
||
Temperature performance index: |
|||
Temperature range |
A:-40℃~+150℃ |
B:-20℃~+150℃ |
D:-70℃~+150℃ |
Temperature uniformity |
≤1.0℃(-40℃~+100℃/-70℃~+100℃); ≤1.5℃(-40℃~+150℃/-70℃~+150℃) |
||
Temperature deviation |
≤±1.0℃(-40℃~+100℃/-70℃~+100℃); ≤±1.5℃(-40℃~+150℃/-70℃~+150℃) |
||
Temperature fluctuation |
≤±0.3℃ (≤±0.5℃ Base on GB/T5170-1996 Standard) |
||
Heating rate |
-40℃ (-70℃) →+150℃ heating rate: 1~6℃/min nonlinear (average) adjustable; +25℃→+150℃ heating rate: 1~5℃/min nonlinear (average) adjustable; +25℃→+150℃ heating rate: 0.1~3℃/min linear slope adjustable; (Temperature rate 4 ~ 25℃/ min Linear/nonlinear for special conditions need to be customized). |
||
Cooling rate |
+150℃→-40℃ Cooling rate: 1~3℃/min nonlinear (average) adjustable; +150℃→-70℃ Cooling rate: 1~2.5℃/min nonlinear (average) adjustable; +25℃→-40℃ (-70℃) Cooling rate: 1~1.7℃/min nonlinear (average) adjustable; +25℃→-40℃ (-70℃) Cooling rate: 1℃/min linear slope adjustable; (Temperature rate 2 ~ 25℃/ min Linear/nonlinear for special conditions need to be customized |
||
Humidity performance index: |
|||
Temperature range |
+10℃~+95℃。 |
||
Humidity range |
10%RH~98%RH(5%~98%R.H;Customized for special conditions) |
||
Humidity uniformity |
≤3.0%RH。 |
||
Humidity deviation |
+2.0%RH~-3.0%RH(±3.0%RH Base on GB/T 2423.3-2008 Standard) |
||
Humidity fluctuation |
≤±2.0%RH |
||
Temperature and humidity all test points |
For details, see 2.2 Temperature and Humidity Control Capability Diagram. |
||
Time resolution of temperature and humidity |
Temperature:0.01℃;Humidity: 0.1%;Time: 0.01min |
||
Standard configuration |
observation Windows (double-layer hollow tempered conductive glass) 1pcs, test holes Φ50mm/Φ100mm (located on the left) 1pcs, metal net sample rack 2pcs, LED lights 1pcs in the box, water supply box 1pcs, weather belts(for wet bulb) 1pcs, small casters 4pcs, power cord 1pcs |